Electronic SystemS TF-108 Manuale dell'Operatore Pagina 6

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 31
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 5
Operator's Manual
6
TF-DSQ-OM-E RevC
TF-DSQ Overview
Probe Calibration with the TF-DSQ Fixture
The TF-DSQ fixture is used in conjunction with the oscilloscope software to perform probe
deskew and DC calibration. It has the following leading features and specifications:
Deskew to +/- 20 ps typical accuracy
Differential and single-ended drive
75 ps edge
Calibration of DC gain and offset and skew at same probing point
Accounts for risetime variations
Accounts for common-mode voltage
DC gain calibration accounts for probe loading effects
Integrated operation with oscilloscope for fully automatic calibration
WavePro 7 Zi and
WaveMaster 8 Zi
Oscilloscopes and SDA/DDA
7 and 8 Zi Models
WavePro 7000
Oscilloscopes
WaveMaster 8000
Oscilloscopes and SDA/DDA
Models, also DDA 5005A,
5005A XXL
Oscilloscope
Connection
ProLink and ProBus
ProBus Only
ProLink Only
Probes
Supported
D610,D620, D600A-AT, D310, D320, and D300A-AT with all probing accessories
AP020, AP033, AP034 HFP1000, 1500, 2500, and 3500, PP005 and PP005A
DC Range
±5 V single-ended, ±10 V differential
DC Accuracy
±(1% + 600 μV)
Edge Risetime
(of Device)
75 ps (typical) < 95 ps (guaranteed)
Edge Amplitude
and Rep Rate
Approximately 800 mV @ 10 MHz
Deskew
Accuracy
±20 ps (typical)
Vedere la pagina 5
1 2 3 4 5 6 7 8 9 10 11 ... 30 31

Commenti su questo manuale

Nessun commento